A total internal reflection fluorescence microscope (TIRFM) is a type of microscope with which a thin region of a specimen, usually less than 200 nanometers can be observed.

TIRFM is an imaging modality which uses the excitation of fluorescent cells in a thin optical specimen section that is supported on a glass slide. The technique is based on the principle that when excitation light is totally internally reflected in a transparent solid coverglass at its interface with a liquid medium, an electromagnetic field, also known as an evanescent wave, is generated at the solid-liquid interface with the same frequency as the excitation light. The intensity of the evanescent wave exponentially decays with distance from the surface of the solid so that only fluorescent molecules within a few hundred nanometers of the solid are efficiently excited. Two-dimensional images of the fluorescence can then be obtained, although there are also mechanisms in which three-dimensional information on the location of vesicles or structures in cells can be obtained.

History Widefield fluorescence was introduced in 1910 which was an optical technique that illuminates the entire sample. Confocal microscopy was then introduced in 1960 which decreased the background and exposure time of the sample by directing light to a pinpoint and illuminating cones of light into the sample. In the 1980s, the introduction of TIRFM further decreased background and exposure time by only illuminating the thin section of the sample being examined.

Background There are two common methods for producing the evanescent wave for TIRFM. The first is the prism method which uses a prism to direct the laser toward the interface between the coverglass and the media/cells at an incident angle sufficient to cause total internal reflection. This configuration has been applied to cellular microscopy for over 30 years but has never become a mainstream tool due to several limitations. Although there are many variations of the prism configuration, most restrict access to the specimen which makes it difficult to perform manipulations, inject media into the specimen space, or carry out physiological measurements. Another disadvantage is that in most configurations based on the inverted microscope designs, the illumination is introduced on the specimen side opposite of the objective optics which requires imaging of the evanescent field region through the bulk of the specimen. There is great complexity and precision required in imaging this system which meant that the prism method was not used by many biologists but rather limited to use by physicists.

The other method is known as the objective lens method which has increased the use of TIRFM in cellular microscopy and increased furthermore since a commercial solution became available. In this mechanism, one can easily switch between standard widefield fluorescence and TIRF by changing the off-axis position of the beam focus at the objective's back focal plane. There are several developed ways to change the positions of the beam such as using an actuator that can change the position in relation to the fluorescence illuminator that is attached to the microscope.

Application In cell and molecular biology, a large number of molecular events in cellular surfaces such as cell adhesion, binding of cells by hormones, secretion of neurotransmitters, and membrane dynamics have been studied with conventional fluorescence microscopes. However, fluorophores that are bound to the specimen surface and those in the surrounding medium exist in an equilibrium state. When these molecules are excited and detected with a conventional fluorescence microscope, the resulting fluorescence from those fluorophores bound to the surface is often overwhelmed by the background fluorescence due to the much larger population of non-bound molecules. TIRFM allows for selective excitation of the surface-bound fluorophores, while non-bound molecules are not excited and do not fluoresce. Due to the fact of sub-micron surface selectivity, TIRFM has become a method of choice for single molecule detection.

There are many applications of TIRFM in cellular microscopy. Some of these applications include:

Measuring the kinetics of receptor endocytosis in response to ligand binding and receptor movement Observing exocytic events through the loading of vesicles undergoing exocytosis with fluorescent dyes Qualitatively and quantitatively describing the roles different proteins play in endocytosis/exocytosis Observing the size, movement, and distance apart of the regions of contact between a cell and a solid substrate

With the ability to resolve individual vesicles optically and follow the dynamics of their interactions directly, TIRFM provides the capability to study the vast number of proteins involved in neurobiological processes in a manner that was not possible before.

Benefits TIRFM provides several benefits over standard widefield and confocal fluorescence microscopy such as:

The background is substantially decreased so structures can be seen clearly There is virtually no out-of-focus fluorescence collected which decrease blurring effects Cells are exposed to a significantly smaller amount of light which limits phototoxicity to cells

Overview The idea of using total internal reflection to illuminate cells contacting the surface of glass was first described by E.J. Ambrose in 1956. This idea was then extended by Daniel Axelrod at the University of Michigan, Ann Arbor in the early 1980s as TIRFM. A TIRFM uses an evanescent wave to selectively illuminate and excite fluorophores in a restricted region of the specimen immediately adjacent to the glass-water interface. The evanescent electromagnetic field decays exponentially from the interface, and thus penetrates to a depth of only approximately 100 nm into the sample medium. Thus the TIRFM enables a selective visualization of surface regions such as the basal plasma membrane (which are about 7.5 nm thick) of cells. Note, however, that the region visualized is at least a few hundred nanometers wide, so the cytoplasmic zone immediately beneath the plasma membrane is necessarily visualized in addition to the plasma membrane during TIRF microscopy. The selective visualization of the plasma membrane renders the features and events on the plasma membrane in living cells with high axial resolution.

TIRF can also be used to observe the fluorescence of a single molecule,Andre et al. Cross-correlated tirf/afm reveals asymmetric distribution of forcegenerating heads along self-assembled, synthetic myosin filaments. Biophysical Journal, 96:1952–1960, 2009. making it an important tool of biophysics and quantitative biology. TIRF microscopy has also been applied in the single molecule detection of DNA biomarkers and SNP discrimination.

Cis-geometry (through-objective TIRFM) and trans-geometry (prism- and lightguide based TIRFM) have been shown to provide different quality of the effect of total internal reflection. In the case of trans-geometry, the excitation lightpath and the emission channel are separated, while in the case of objective-type TIRFM they share the objective and other optical elements of the microscope. Prism-based geometry was shown to generate clean evanescent wave, which exponential decay is close to theoretically predicted function. In the case of objective-based TIRFM, however, the evanescent wave is contaminated with intense stray light. The intensity of stray light was shown to amount 10–15% of the evanescent wave, which makes it difficult to interpret data obtained by objective-type TIRFM

Mechanism

The basic components of TIRFM devices are in the figure to the right.

Objective-based vs prism-based Key differences between objective-based (cis) and prism-based (trans) TIRFM are that prism based TIRFM requires usage of a prism/solution interface to generate the evanescent field, while objective-based TIRFM does not require a prism and utilizes a cover slip/solution interface to generate the evanescent field. Typically objective-based TIRFM are more popularly used, however have lowered imaging quality due to stray light noise within the evanescent wave.

Prism-based

Less extraneous scattering Much cheaper (hundreds instead of thousands of dollars) Best for low-mid range magnification and water immersion objectives Easiest with free collimated laser sources Larger range of incidence angles Desirable to achieve smallest evanescent field depth

Objective-based

High magnification and aperture Stable, easy to set up and align Works with free collimated laser, optical fiber, or conventional arc sources

Methodology

Fundamental physics TIRFM is predicated on the optical phenomena of total internal reflection, in which waves arriving at a medium interface do not transmit into medium 2 but are completely reflected back into medium 1. Total internal reflection requires medium 2 to have a lower refractive index than medium 1, and for the waves must be incident at sufficiently oblique angles on the interface. An observed phenomena accompanying total internal reflection is the evanescent wave, which spatially extends away perpendicularly from the interface into medium 2, and decays exponentially, as a factor of wavelength, refractive index, and incident angle. It is the evanescent wave which is used to achieve increased excitation of the fluorophores close to the surface of the sample, and diminished excitation of superfluous fluorophores within solution.

For practical purposes, in objective based TIRF, medium 1 is typically a high refractive index glass coverslip, and medium 2 is the sample in solution with a lower refractive index. There may be immersion oil between the lens and the glass coverslip to prevent significant refraction through air.

Evanescent wave The critical angle for excitatory light incidence can be derived from Snell's law:

For the refractive index of sample, the refractive index of the cover slip.

Thus, as the angle of incidence reaches , we begin observing effects of total internal reflection and evanescent wave, and as it surpasses these effects are more prevalent.

The intensity of the evanescent wave is given by:

With penetration depth given by:

Typically, ≤~100 nanometers, which is typically much smaller than the wavelength of light, and much thinner than a slice from confocal microscopes.

For TIRFM imaging the wavelength of the excitation beam within the sample can be selected for by filtering. Additionally, the range of incident angles is determined by the numerical aperture (NA) of the objective, and requires that NA > . This parameter can be adjusted by changing the angle the excitation beam enters the objective lens. Finally, the reflective indices () of the solution and cover slip can be experimentally found or reported by manufacturers.

Excitation beam For complex fluoroscope microscopy techniques, lasers are the preferred light source as they are highly uniform, intense, and near-monochromatic. However, it is noted that ARC LAMP light sources and other types of sources may also work. Typically the wavelength of excitation beam is designated by the requirements of the fluorophores within the sample, with most common excitation wavelengths being in the 400–700 nm range for biological samples.

In practice, a lightbox will generate a high intensity multichromatic laser, which will then be filtered to allow the desired wavelengths through to excite the sample. For objective-based TIRFM, the excitation beam and fluoresced emission beam will be captured via the same objective lens. Thus, to split the beams, a dichromatic mirror is used to reflect the incoming excitation beam towards the objective lens, and allow the emission beam to pass through into the detector. Additional filtering may be required to further separate emission and excitation wavelengths.

Emission beam